Where does meter accuracy variation come from ...
Machining tolerances:
Variation within a production run is normally minimal. The run is sized so as
to not allow tool wear to shift the tolerances. In the case of the C700 cold
water meter, the inside of the working chamber is machined in two passes. A variation
of about 2/10,000" on the depth of the cut across the bottom, will shift
the accuracy about 0.1%. In general the lower the viscosity of the fluid, the
greater the accuracy loss with this type of machining variation.
You can see machining error in the working chamber, it will appear as a slight
line in the center of the space between the hub and the inside wall. You will
probably never see this in a AMCO water meter.
Temperature variations:
Meter tolerances are designed for "ambient" operation. On the AMCO
oil meters, with temperature rises the anodized aluminum piston and the brass
working
chamber will expand at different rates, varying the gap between the piston OD
and the chamber ID. This will change the amount of leakage through the meter.
Testing technique:
Testing technique, especially on oil meters is important. The test container
should drain fully, or at least be positioned in a immovable position so as to
get repeatability of any trapped fluid after each draindown period. One should
always fill the test container and drain it for a preset period of time. This
will provide a uniform oil film on the inside tank walls (technical term is "clingage").
With a nonuniform film from run to run, the operator will observe apparent meter
nonrepeatability and accuracy error. On heavier oils film thickness variation
can severely distort test results.
Flow rate adjustment during the test cycle ... an operator will tend to slow
down the flow rate at the end of the test in order to meter out a known volume
precisely. He will frequently drop the flow below the minimum flow rate of the
meter, and introduce a serious error for that period.
Gravimetric testing:
Testing by weight is one of the best ways to evaluate a meters accuracy. However,
if a movable platform scale, either mechanical or electronic, is used ... has
it been banged around, is it still accurate?
Typically, a customer will place a 55 gallon drum on a weighscale, take a tare
weight, fill the tank and convert the weight to gallons to check the meter. Assume
that he runs 50 gallons (417.26 pounds) of water. To this 417.26 pounds you add
the tare weight of the drum and arrive at something on the order of 450-460 pounds.
An error of 1/2% percent on the meter is 1/4 gallon or 2.08 pounds. The question
is whether the customer can read a tare weight at 30-40 pounds and a total weight
of 450-460 pounds, subtract the two readings and have enough accuracy in his
two measurements to properly judge the flow meters performance.
This reading resolution and accuracy is possible with an electronic scale with
load cells, but can be very questionable on a mechanical scale. People kid themselves
about the accuracy of their scales.
Plastic bodied meters:
If the customer uses a clamping device which exerts a compressive force on the
meter ends, it is possible to distort the body shape, and create a gap between
the fit of the working chamber and the body. This allows leakage around the working
chamber, and creates a "low" reading on the register.
The best way:
The best measuring system uses a electronic gravimetric (weighscale) scale with
a flow diverter. You run the water through the flow meter at a preset rate and
as the pointer crosses a cardinal point you hit a button. This button actuates
a cylinder which diverts the flow into the weigh tank. When the pointer passes
another cardinal point you hit another button and the diverter flips back to
the original position. The result is a uniform rate of flow test with high accuracy.
In a production situation, one would normally use a photosensor which detects
the pointer position, thus eliminating human variation and reaction time.
General ... Comparison test meter
Turbine meters such as the Sponsler Precision turbine meter are typically used
as comparison test meters. They come with 10-point calibrations with NIST traceability,
are 1/2% (or better), they are small, portable and easily stored. However, you
have to treat them properly:
A comparison standard meter which is removed from the flow line and only used
for periodic testing requires flushing after each use. To eliminate the small
deposits which form when water is evaporated, the meter should be flushed with
alcohol and allowed to air dry before storage. We recommend the use of 95% denatured
ethyl alcohol which is commercially sold as "stove alcohol". As a alternative,
91% isopropyl alcohol containing less than 0.003% by weight non-volatile material
can be used. Do not use wood alcohol (methanol), rubbing alcohol, or ethyl alcohol
with non-volatile denaturing additives. The flow meter is a precision turbine
and while it can tolerate some abuse, to avoid a shift in accuracy or response
due to buildup of foreign matter in the bearing structure, proper care prior
to storage is recommended.
To obtain the full accuracy potential of the turbine meter, it must be installed
in a laminar flow situation in horizontal piping. This means at least 10 horizontal
pipe diameters of straight pipe ahead and 5 horizontal diameters after the meter.
General ... Installation notes and commentary
If the flowmeter is a turbine design, vortex shedding, paddlewheel:
Allow as much straight pipe, in the same size as the flowmeter line size, before
and after the meter. The minimum should be 10 pipe diameters ahead and 5 pipe
diameters after the meter. If you have a 4" meter, this means 40" ahead
of the meter and 20" after the meter.
If the flowmeter is a positive displacement design, straight pipe is not
an important issue.
Some general comments:
a. Flowmeters do not work well with silty or dirty fluids. Use strainers ahead
of the meter if this is your situation.
b. Teflon tape, pipe dope ... do not, under any circumstances, get tape or pipe
dope over the end of the connecting pipe. When screwed in, it will break free
and wash into the flowmeter. This means trouble.
c. Does the meter match the application ...
... if you have hot water, is the meter rated for it?
... if you have chemicals, can the meter handle the fluid?
... is the line full? The meter must not have any air in it.
... if a gravity line installation, is there enough head pressure?
... think about your installation, what can create a problem?
d. After installation, don't open a valve and have water come rushing down the
pipe at a high rate. The escaping air will cause the meter element to spin rapidly
and in extreme cases self destruct. When the water reaches the meter a high velocity
wall of water can damage the meter when it hits the working element.
e. Water hammer. This is what you hear in a pipe which has a high flow rate going
through it and you slam a valve closed quickly. Moving water has a lot of energy
in it. If you have a 2" pipe system with about 100 gallons of water distributed
through it, and it is flowing at 200 GPM, this moving water looks like 830 pounds
moving at 13 MPH. When that valve slams closed, you are asking it to stop that
830 pounds instantaneously. The bang you hear is a shock wave. This shock wave
can make pipes split open. Avoid water hammer by using slow closing valves, or
valves which close in 2-stages, reducing the water flow before doing the final
closing.
|